Bruker 900-MHz NMR (CSB)
The Bruker 900-MHz Avance II spectrometer has a cryoprobe TXI probe with single, axis gradients, a room temperature TCI probe with triple axis gradients and temperature controller. The instrument is located at the Center for Structural Biology at the University of Illinois at Chicago.
Specifications
Part | Description |
---|---|
Magnet | Bruker 900-MHz US² |
Probes | 1) TCI Cryoprobe with ¹H/²H/¹³C/¹⁵N pulsing and z-axis gradients |
2) TXI probe ¹H/²H/¹³C/¹⁵N with 3-axis gradients | |
Console | Avance II |
Software | Topspin v1.3 |
Nutation Field Strengths
Nucleus | Power Level | pw90 | Notes |
---|---|---|---|
1H (900.1MHz) | -1.0 dB | 12μs1 | Hard pulse (<100μs) |
13C (226.3MHz) | 1.2 dB | 15μs | Hard pulse (<100μs) |
13.2 dB | 60.0μs | Decoupling pulse (<80ms) | |
15N (91.2MHz) | -0.8 dB | 45μs | Hard pulse (<100μs) |
12.1 dB | 200μs | Decoupling pulse (<80ms) |
-
Calibrate on every sample ↩